型號 Type
|
HX-1000TM/LCD
|
HXD-1000TM/LCD
|
物鏡/壓頭切換
Lens/Indenter switch
|
自動完成物鏡與壓頭轉塔過程
With Motorized Turret
|
轉塔定位誤差
Position accuracy
|
2μm以內
Not greater than 2μm
|
試驗力
Test force
|
0.0981,0.2452,0.4903,0.9807,1.961,2.942,4.903,9.807 ( N)
10, 25, 50, 100, 200, 300, 500, 1000 (gf)
|
加荷方式
Loading mode
|
程控自動加荷,保荷和卸荷
Program-controlled application, duration and removal of load
|
保荷時間
Duration time
|
5–60 sec selectable 步長 (In step of ) 5 sec.
|
硬度測量范圍
Measuring range
|
5 HV – 3000 HV以上above
硬度示值誤差Indication error≤3%
|
總放大倍率
Total Magnification
|
物鏡 Objective: 40x; 10x(觀察用 For observation),
測量目鏡 Eyepiece: 15x (可選配Optional :10X)
總倍率 Total: 600x; 150x (觀察用 For observation)
|
最小計數單位 |
0.01μm
|
最小檢測單位
Min. detection
|
0.025μm
|
X-Y測試臺
X-Y stage
|
尺寸 Dimensions:100×100mm 最大移動 Travel range:25×25 mm
最小讀數Resolution:1/100mm
|
光路切換
Optical path
|
雙光通道(測量/攝影切換)
Measurement / Photo switch
|
測量讀數方式
Measuring mode
|
用測微目鏡取得對角線長度,輸入,即在LCD顯示硬度值,可打印輸出. After taking D1(D2) with the eyepiece and entering them, the hardness value is displayed on LCD. The data can be printed out. |
目鏡測量數字顯示, 數字顯示試驗力,保荷時間, 硬度值, 最大最小值, 平均值, 不均勻度, 標準偏差, 洛氏HRC HRA轉換值等。內置打印機提供測試報告. With eyepiece digital display, Display on LCD: Load, Duration time, Diagonal lengths, Hardness values, Min, Max, Average, Divergence,, Standard deviations, converted HRC HRA values, and etc. A test result is provided with the built-in printer. |
數據輸出
Data output
|
RS-232接口
RS-232 interface
|
試件最大高度
Max..test height
|
85mm/90mm
|
執(zhí)行標準(或參照)
Standards concerned
|
中國國家標準 GB/T 4340
國際標準ISO 6507,美國標準ASTM E384,日本標準 JIS B7734
|